New TopShim run:

version = topshim 4.4.1
date = 2025-09-28
time = 12:59:22.806
command = topshim 

FINE SHIMMING

Parameters:

probe = Z114607_0305
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration 1: FWHM Before= 0.2 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration 2: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration 3: FWHM Before= 0.5 Hz, Target= 0.3 Hz, After= 0.3 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.446

shim changes: 
Z	-87
Z2	-134
Z3	-98
Z4	-660
Z5	-48

duration = 54 sec
completed successfully